Applied Biosystems™ ViiA™ 7 Array Card RNaseP Verification Kit
Used to validate the performance of the ViiA™ 7 Real-Time PCR System with the TaqMan™ Low Density Array Block Upgrade.
Brand: Applied Biosystems™ 4432265
1705.00 EUR valid until 2024-12-20
Use promo code "21651" to get your promotional price.
Includes: An empty TaqMan Low Density Array and eight vials of RNase P detection reagents to detect and quantitate genomic copies of the human RNase P gene. Each vial contains premixed reagents (1X TaqMan Universal PCR Master Mix, RNase P primers, and FAM™ dye-labeled probe) and template.
Description
The ViiA™ 7 System TaqMan™ Low Density Array Instrument Verification RNase P kit contains an empty TaqMan™ Low Density Array and 8 vials of RNase P detection reagents to detect and quantitate genomic copies of the human RNase P gene, a single-copy gene encoding the RNA moiety of the RNase P enzyme. Each vial contains premixed reagents (1X TaqMan™ Universal PCR Master Mix, RNase P primers, and FAM™ dye-labeled probe) and template. The ViiA™ 7 System TaqMan™ Low Density Array Instrument Verification RNase P kit must demonstrate the ability to distinguish between 5,000 and 10,000 genomic equivalents with a 99.7% confidence level for a subsequent sample run in a single well. To ensure accurate results, Applied Biosystems™ recommends that the performance of your instrument be verified every six months.
The kit must demonstrate the ability to distinguish between 5000 and 10,000 genomic equivalents with a 99.7% confidence level for a subsequent sample run in a single well. To ensure accurate results, the instrument should be verified every six months.
Order Info
Store kit at -15° to -25°C.;
Specifications
384-well Microfluidic Card | |
ViiA™ 7 System | |
Array Card RNaseP Verification Kit | |
1 kit |
8 x Tubes with Standard Dilutions, Empty Array Card, Port 1 NTC, Port 2 Unknown A, Port 3 U, Store at -15°C to -25°C | |
ViiA™ | |
Real Time PCR (qPCR) | |
RNase P Calibration, Instrument Verification Plate |
For Research Use Only. Not for use in diagnostic procedures.